Implanted Components

 

  • The VNS Therapy™ System has been implanted in more than 45,000 patients with epilepsy
  • Back-up of all implanted components are needed at all surgeries


VNS Therapy Generators

  • The VNS Therapy generator is an implantable, programmable pulse generator that delivers a precise pattern of stimulation to the left vagus nerve.


Demipulse™ Model 103 Generator

  • For new patients or replacement with a single pin lead

Physical Characteristics of the Model 103 Generator
Titanium Case with Header 6.9 mm x 45 mm x 32 mm
Weight 16 grams
Header Polyurethane, single pin receptacle
Lead Compatibility Model 302 or Model 303 (single pin)
Power Source Lithium Carbon Monofluoride Battery
Average Expected Longevity ~6+ years depending on stimulation parameters*


*Predicted longevity at 2 mA, 20 Hz, 500 msec, 10% duty cycle into a 4 kOhm load. The dose settings affect how long the battery will last. For example, the battery may last for 3 years at a higher setting compared with 8 years at a lower setting.


Demipulse™ Model 104 Generator

  • For replacement with a dual pin lead

Physical Characteristics of the Model 104 Generator
Titanium Case with Header 6.9 mm x 45 mm x 39 mm
Weight 17 grams
Header Polyurethane, dual pin receptacle
Lead Compatibility Model 300 (dual pin)
Power Source Lithium Carbon Monofluoride Battery
Average Expected Longevity ~6+ years depending on stimulation parameters*


*Predicted longevity at 2 mA, 20 Hz, 500 msec, 10% duty cycle into a 4 kOhm load. The dose settings affect how long the battery will last. For example, the battery may last for 3 years at a higher setting compared with 8 years at a lower setting.


Pulse™ Model 102 Generator

  • For new patients or replacement with a single pin lead

Physical Characteristics of the Model 102 Generator
Titanium Case with Header 6.9 mm x 52.2 mm x 51.6 mm
Weight 25 grams
Header Polyurethane, single pin receptacle
Lead Compatibility Model 302 or Model 303 (single pin)
Power Source Lithium Carbon Monofluoride Battery
Average Expected Longevity ~6+ years depending on stimulation parameters*


*Predicted longevity at 2 mA, 20 Hz, 500 msec, 10% duty cycle into a 4 kOhm load. The dose settings affect how long the battery will last. For example, the battery may last for 3 years at a higher setting compared with 8 years at a lower setting. For full range of settings in relationship to battery life, see the Physician’s Manual or the EOS Projection on the handheld.


Pulse Duo™ Model 102R Generator

  • For replacement with dual pin lead

Physical Characteristics of the Model 102R Generator
Titanium Case with Header 6.9 mm x 58.9 mm x 51.6 mm
Weight 27 grams
Header Polyurethane, dual pin receptacle
Lead Compatibility Model 300 (dual pin)
Power Source Lithium Carbon Monofluoride Battery
Longevity ~6+ years depending on stimulation parameters*


*Predicted longevity at 2 mA, 20 Hz, 500 msec, 10% duty cycle into a 4 kOhm load. The dose settings affect how long the battery will last. For example, the battery may last for 3 years at a higher setting compared with 8 years at a lower setting. For full range of settings in relationship to battery life, see the Physician’s Manual or the EOS Projection on the handheld.


The Model 100 and Model 101 Generators are no longer distributed


VNS Therapy Leads

  • The VNS Therapy lead is a bipolar electrical lead that transmits stimulation from the VNS Therapy generator to the left vagus nerve in the neck.
  • The lead is made up of the pin that connects to the generator on one end, and the helices that contain the stimulation electrodes and anchor tether on the other end.
  • All lead models are currently available in two sizes based on helical inner diameter:
    • Model 302-20 (2.0 mm)
    • Model 302-30 (3.0 mm)
    • Model 303-20 (2.0 mm)
    • Model 303-30 (3.0 mm)


Simulation of the electrodes and anchor tether attachment to the vagus nerve.


Perennia™ Model 303 Lead

The Model 303 Lead is a single-use-only device intended for use with the Pulse Model 102 or Model 103 generator and is subject to the same indications for use.

Physical Characteristics of the Perennia™ Model 303 Lead (single pin)
Insulation Silicone
Length 43 cm
Resistance 180 to 250 ohms, pin to electrode
Electrode Sizes 2.0 and 3.0 mm inner diameter
Generator Compatibility Model 102 and Model 103 (single pin receptacle)
Conductor Material Platinum Iridium

 


Model 302 Lead

The Model 302 lead is a single-use-only device intended for use with the Pulse Model 102 and Model 103 Generator and is subject to the same indications for use.

Physical Characteristics of the Model 302 Lead (single pin)
Insulation Silicone
Length 43 cm
Resistance 120 to 180 ohms, pin to electrode
Electrode Sizes 2.0 and 3.0 mm inner diameter
Generator Compatibility Model 102 and Model 103 (single pin receptacle)
Conductor Material Platinum Iridium

 


The Model 300 Lead is no longer distributed



Back to Components

 
Indications For Use
Terms Of Use
Privacy Statement
International Site